Three-dimensional line-field Fourier domain optical coherence tomography for in vivo dermatological investigation
Yoshiaki Yasuno1, Takashi Endo, Shuichi Makita
1University of Tsukuba, Institute of Applied Physics, Tennodai 1-1-1, Tsukuba, Ibaraki 305-8573, Japan. yasuno@optlab2.bk.tsukuba.ac.jp
Abstract:
We demonstrate 3-D optical coherence tomography using only 1-D mechanical scanning. This system uses the principle of Fourier domain optical coherence tomography for depth resolution, 1-D imaging for lateral vertical resolution, and mechanical scanning by a galvanometer for lateral horizontal resolution. An in vivo human fingerpad is investigated in three dimensions with an image size of 480 points (vertical) x 300 points (horizontal) x 1024 points (depth), which corresponds to 2.1 x 1.4 x 1.3 mm. The acquisition time for a single cross section is 1 ms and that for a single volume is 10 s. The system sensitivity is 75.6 dB at a probe beam power of 1.1 mW.