X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Optimisation of imaging technique used in direct digital radiography
J A Roberts1, S C Evans, M Rees
1Medical Physics and Clinical Engineering, Singleton Hospital, Swansea SA2 8QA, UK.
Abstract:
The purpose of the study was to optimise the technique employed for AP shoulder and lateral cervical spine examinations following an investigation into image quality, based on clinical assessment, and effective dose, calculated from patient entrance surface dose measurements. A study was therefore conducted in an attempt to determine whether the increased radiation dose to the patient following the introduction of an anti-scatter grid was justified by the level of improvement in image quality. The study, involving 100 patients, was able to demonstrate that the increase in radiation dose to the patient when using an anti-scatter grid for AP shoulder examinations is not justified by the improved image quality. A poor level of inter-rater reliability between the consultants scoring the lateral cervical spine images prevented a firm conclusion from being reached. The fact that all images were of diagnostic quality, however, suggested that the use of the anti-scatter grid was unnecessary. Following completion of the project the hospital involved was informed of all findings.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...