Pacemaker mediated tachycardias in single chamber rate responsive pacing.
Pacing and Clinical Electrophysiology : PACE
|December 1, 1990
Summary
Pacemaker flipping in single-chamber devices can cause pacemaker-mediated tachycardia. Repositioning the pacemaker or reducing output can resolve this issue, preventing further complications.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Pacemaker-mediated tachycardia (PMT) is typically linked to dual-chamber pacemakers.
- Single-chamber, rate-responsive pacemakers can also develop PMT under specific conditions.
Observation:
- An 83-year-old man with complete atrioventricular block developed palpitations and pocket twitching post-implantation of a unipolar activity-sensing pacemaker.
- The pacemaker had flipped, causing diaphragmatic pacing that activated the rate-responsive sensor, initiating a positive feedback loop and tachycardia.
Findings:
- Pacemaker-mediated tachycardia occurred due to pacemaker flipping and subsequent pocket pacing activating the rate-responsive sensor.
- Reducing pacemaker output to 2.5 V successfully eliminated pocket pacing and the tachycardia.
- Pacemaker reversal necessitated a more sensitive setting for achieving adequate rate response during exercise.
Implications:
- Pacemaker flipping is a potential complication leading to PMT in single-chamber activity-sensing pacemakers.
- Strategies for preventing and managing pacemaker flipping are crucial.
- Understanding sensor-mediated tachycardias is important for device management.