Comparison of scatter doses from a multislice and a single slice CT scanner
1Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia. John.Burrage@health.wa.gov.au
Manufacturer data for multislice CT (MSCT) scanners overestimated scatter doses. Actual measurements showed MSCT scatter doses were approximately 4 times higher than single slice CT (SSCT), aligning with beam width differences and confirming existing shielding adequacy.
Area of Science:
- Medical Physics
- Radiological Sciences
- Imaging Technology
Background:
- Manufacturer data suggested significantly higher external scatter doses for multislice CT (MSCT) compared to single slice CT (SSCT).
- This discrepancy contradicted expectations and UK ImPACT group statements regarding scatter dose equivalence when accounting for beam width.
Purpose of the Study:
- To investigate and compare external scatter doses between MSCT and SSCT scanners.
- To validate manufacturer's scatter dose data and assess shielding adequacy for a new MSCT installation.
Main Methods:
- Scatter dose measurements were conducted at three points using a standard perspex CTDI phantom for both MSCT (40 mm beam width) and SSCT (10 mm beam width).
- CT dose indices were measured to compare scanner outputs.
- A film badge survey was performed post-installation to evaluate room shielding effectiveness.
Main Results:
- Measured MSCT scatter doses were lower than manufacturer's predictions.
- MSCT scatter doses were approximately 4 times higher than SSCT doses, consistent with the difference in beam widths.
- CT dose indices were comparable between the two scanner types.
- The existing lead shielding was found to be adequate for the MSCT scanner.
Conclusions:
- Manufacturer's data for MSCT external scatter doses may be overestimated.
- Measured scatter dose increases with MSCT are attributable to wider beam widths, not disproportionately higher scatter.
- Existing shielding protocols are sufficient for MSCT, provided they were adequate for SSCT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT