Related Experiment Video
Updated: Jul 13, 2026

In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
[Improved visualization of foveal pathologies using fourier-domain optical coherence tomography]
Masanori Hangai1, Yumiko Ojima, Akiko Yoshida
1Department of Ophthalmology and Visual Science, Kyoto University Graduate School of Medicine, Japan. hangai@kuhp.kyoto-u.ac.jp
Purpose:
To demonstrate the improved visualization of pathological foveal structures by a prototytpe Fourier-domain optical coherence tomography instrument (FD-OCT).
Materials And Methods:
We constructed a prototype FD-OCT instrument with an axial resultion of 4.3, m in tissues and an acquisition rate of approximately 18,700 axial scans per second. A raster-scan protocol comprising 256 x 256 axial scans was used in a 2.8 x 2.8 mm area to obtain a 3D-OCT data set. Images of the fovea were obtained in patients with central serous chorioretinopathy(5 eyes of 5 patients), macular hole (6 eyes of 5 patients), pigment epithelial detachment (PED : 4 eyes of 4 patients), and epiretinal membrane (4 eyes of 4 patients).
Results:
Normal fovea showed two lines corresponding to a strong backreflection from the external limiting membrane (ELM) and the inner and outer segment junctions (IS/OS). With the exception of the eyes with epiretinal membrane, the IS/ OS line that corresponded to the region where the photoreceptor layer shifted inward disappeared in all eyes. The ELM lines were clearly observed in the same region, except for the one eye with the most steeply protruding PED.
Conclusions:
The use of FD-OCT enables improved visualization of the pathological structural changes in the foveal photoreceptor layer of patients with macular diseases.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Glaucoma: Overview
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Imaging Studies VII: Vascular Imaging