[Nimesulide spectrophotometric determination in the visible region]
Andreia Claudia Perju1, Mariana Mândrescu, A F Spac
1Facultatea de Farmacie, Disciplina de Chimie analitică, Universitatea de Medicină i Farmacie "Gr.T. Popa" Iaşi.
Unlabelled:
Nimesulide is a nonsteroidal antiinflammatory (NSAID) drug whose mechanism of action is characterized by selective inhibition of cyclooxygenase-2. Nimesulide have analgesic and antipyretic properties.
Materials And Method:
For nimesulide analysis we developed a spectrophotometric method based on reduction of the nitro group of nimesulide by zinc and hydrochloric acid followed by diazotization and coupling with thymol (5-methyl-2-(1-methylethyl) phenol) in ammonia medium to form a stable chromophore which absorbs at 476 nm. There were established the practical working conditions. The developed method was validated.
Results:
The method showed a good linearity in the range of 5 - 40 microg / sample (the correlation coefficient r = 0.9990). The detection limit (LD) was 0.99 microg/sample and the quantification limit (LQ) was 3.32 microg/sample. There were established the system precision (RSD = 1.70%), the precision (RSD < 2.2%) and the accuracy (recovery in the range 97.66 - 102.69% with a mean recovery of 99.81%) of the method.
Conclusions:
A new method for nimesulide spectrophotometric quantification in visible has been developed.
More Related Videos
Related Concept Videos
UV–Vis Spectrometers
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Amines to Sulfonamides: The Hinsberg Test
Generally, a primary amine reacts with the Hinsberg reagent to produce an N-substituted benzenesulfonamide. The electron-withdrawing sulfonyl...
IR Frequency Region: Fingerprint Region
The...
UV–Vis Spectrum
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar absorptivity (ε) or log ε on the y-axis (ordinate)...