Determinants of gradient field-induced current in a pacemaker lead system in a magnetic resonance imaging environment
Harikrishna Tandri1, Menekhem M Zviman, Steven R Wedan
1Johns Hopkins University School of Medicine, Baltimore, MD 21287, USA. htandri1@jhmi.edu <htandri1@jhmi.edu>
Background:
The determinants of low-frequency-induced current by magnetic resonance imaging (MRI) gradient fields in a pacemaker lead system are largely unknown.
Objective:
The purpose of this study was to determine the magnitude of MRI low-frequency-induced current in an implanted pacemaker lead system and to investigate in vivo determinants of low-frequency-induced current in an animal model.
Methods:
Six mongrel dogs underwent conventional single-chamber pacemaker implantation with a current recorder connected in series. Pulse generator (PG) was programmed to VOO 120 bpm with subthreshold output. MRI was performed in a 1.5-T scanner. Low-frequency-induced current was recorded during unipolar pacing, bipolar pacing, and bipolar pacing with the PG case electrically isolated from the pocket. In each mode, low-frequency-induced current was recorded with and without a large loop of additional lead connected in series.
Results:
With a conventional implant, low-frequency-induced current was < or =0.5 mA in all three pacing modes. With five external loops, the magnitude of low-frequency-induced current increased to >30 mA, with consistent myocardial capture in unipolar and bipolar pacing. However, in bipolar pacing with the PG electrically isolated from the pocket, low-frequency-induced current decreased to <0.5 mA with no myocardial capture even with additional looped leads.
Conclusion:
Under conventional implant conditions, the magnitude of low-frequency-induced current is <0.5 mA and is unlikely to cause myocardial capture; however, arrhythmia induction cannot be excluded. With sufficient increase in effective loop area (additional looped leads), direct myocardial capture by the low-frequency-induced current is possible. In this study, breaking the return pathway by electrically isolating the PG case from the circuit abolished low-frequency-induced current.
Related Concept Videos
Magnetic Resonance Imaging
Induced Electric Fields: Applications
Magnetic Force On A Current-Carrying Conductor
Consider a compass placed near a current-carrying wire. The wire experiences a force that aligns the needle of the compass tangentially around the wire. Thus, the current-carrying wire produces concentric circular loops of magnetic field. The magnetic field generated by a wire can be...
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Field Of A Current Loop
Magnetic Field due to Moving Charges
Consider a point charge moving with a constant velocity. Like the electric field, the magnetic field at any point is directly proportional to the magnitude of the charge and inversely proportional to the square of the distance between the source point and the field point. However, unlike the electric field, the magnetic field is always perpendicular to the plane containing the line...