Electromagnetic interference with pacemakers caused by portable media players
Jay P Thaker1, Mehul B Patel, Krit Jongnarangsin
1Division of Cardiovascular Medicine, University of Michigan, Ann Arbor, Michigan, USA.
iPods placed near pacemakers can cause interference. This study found that 51% of patients experienced interference, with iPods emitting signals in the pacemaker frequency range when on and connected to headphones.
Area of Science:
- Biomedical Engineering
- Cardiology
- Electromagnetics
Background:
- Electrical devices generate electromagnetic fields that can interfere with pacemakers.
- Portable media players, like iPods, are increasingly common and often carried near the chest.
Purpose of the Study:
- To investigate whether iPods cause interference with cardiac pacemakers.
Main Methods:
- A prospective, single-blinded study involving 100 pacemaker patients tested with four iPod models.
- Pacemakers were monitored via ECG and programmer telemetry while iPods were placed 2 inches away.
- Testing simulated real-world use with headphones, maximum volume, and no equalizer; electromagnetic emissions were also measured.
Main Results:
- 51% of patients and 20% of tests showed interference; Type I (rate histogram) in 19%, Type II (no rate counter effect) in 32%.
- Interference was reproducible across different pacing configurations, modes, and days.
- Maximum iPod emissions occurred in the pacemaker carrier frequency range when the device was on with headphones attached.
Conclusions:
- iPods positioned within 2 inches of implanted pacemakers can cause significant interference.
- This interference is linked to electromagnetic emissions from the iPods, particularly when in use with headphones.
More Related Videos
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...