Changes in prescribed doses for the Seattle neutron therapy system.
1University of Washington Medical Center, Seattle, WA 98195-6043, USA. alina@u.washington.edu
Improvements to the PRISM treatment planning system enhance neutron dose calculations. Updated algorithms and new data reveal discrepancies, necessitating changes to previously reported neutron therapy prescribed doses.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Neutron therapy utilizes specialized treatment planning systems for dose distribution calculations.
- The PRISM system at the University of Washington Medical Center has historically calculated neutron dose distributions.
- Accuracy in absorbed dose calculations is critical for effective radiation therapy.
Purpose of the Study:
- To enhance the accuracy of neutron absorbed dose calculations.
- To investigate the impact of algorithmic and data improvements in the PRISM treatment planning system.
- To re-evaluate previously reported neutron therapy prescribed doses based on improved calculations.
Main Methods:
- Modified the PRISM algorithm to include an analytical expression for central axis wedge factor dependence on field size and depth.
- Incorporated a newly commissioned set of measured data, enabling the use of wedge profiles at various depths.
- Compared absorbed dose calculations between the original and improved PRISM algorithms.
Main Results:
- The improved PRISM algorithm demonstrated discrepancies compared to older versions, primarily due to accounting for central axis wedge factor dependence and depth-specific wedge profiles.
- Older versions used a constant central axis wedge factor and a single wedge profile, leading to inaccuracies.
- The new algorithm's ability to use depth-dependent wedge profiles significantly impacted dose calculations.
Conclusions:
- The enhancements to the PRISM system have improved the accuracy of neutron absorbed dose calculations.
- Discrepancies identified highlight the limitations of previous calculation methods.
- Previously reported prescribed doses for neutron therapy require revision due to these findings.
Related Concept Videos
Dosage Regimen: Individualization
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Dosage Regimens: Partial Pharmacokinetic Parameters