Samuel Patz1, Iga Muradian, Mirko I Hrovat
1Center for Pulmonary Functional Imaging, 221 Longwood Avenue, Brigham and Women's Hospital, Boston, MA 02115, USA. patz@bwh.harvard.edu
Rationale And Objectives:
Using a novel (129)Xe polarizer with high throughput (1-2 L/hour) and high polarization (approximately 55%), our objective was to demonstrate and characterize human pulmonary applications at 0.2T. Specifically, we investigated the ability of (129)Xe to measure the alveolar surface area per unit volume of gas, S(A)/V(gas).
Materials And Methods:
Variable spin echo time (TE) gradient and radiofrequency (RF) echoes were used to obtain estimates of the lung's contribution to both T(2)* and T(2). Standard multislice ventilation images were obtained and signal-to-noise ratio (SNR) determined. Whole-lung, time-dependent measurements of (129)Xe diffusion from gas to septal tissue were obtained with a chemical shift saturation recovery (CSSR) method. Four healthy subjects were studied, and the Butler et al CSSR formalism (J Phys Condensed Matter 2002; 14:L297-L304) was used to calculate S(A)/V(gas). A single-breath version of the xenon transfer contrast (SB-XTC) method was implemented and used to image (129)Xe diffusion between alveolar gas and septal tissue. A direct comparison of CSSR and SB-XTC was performed.
Results:
T(2)*=135+/-29 ms amd T(2)=326.2+/-9.5 ms. Maximum SNR=36 for ventilation images from inhalation of 1L 86% (129)Xe and voxel volume =0.225 mL. CSSR analysis showed S(A)/V(gas) decreased with increasing lung volume in a manner very similar to that observed from histology measurements; however, the absolute value of S(A)/V(gas) was approximately 40% smaller than histology values. SB-XTC images in different postures demonstrate gravitationally dependent values. Initial comparison of CSSR with XTC showed fairly good agreement with expected ratios.
Conclusions:
Hyperpolarized (129)Xe human imaging and spectroscopy are very promising methods to provide functional information about the lung.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...