Corneal parameters and difference between goldmann applanation tonometry and dynamic contour tonometry in normal eyes
Michele Lanza1, Maria Borrelli, Maddalena De Bernardo
1Eye Department, Centro Grandi Apparecchiature, Second University of Naples, Naples, Italy.
Journal of Glaucoma
|September 17, 2008
Summary
Dynamic contour tonometry (DCT) measures intraocular pressure (IOP) higher than Goldmann applanation tonometry (GAT) in healthy eyes. These IOP measurement devices are not interchangeable and DCT may indicate higher normal IOP values.
Area of Science:
- Ophthalmology
- Biomedical Engineering
Background:
- Intraocular pressure (IOP) is a key metric in glaucoma diagnosis.
- Accurate IOP measurement is crucial for patient management.
- Goldmann applanation tonometry (GAT) is the established standard, but newer devices like dynamic contour tonometry (DCT) offer alternative methods.
Purpose of the Study:
- To compare IOP measurements between DCT and GAT in healthy eyes.
- To investigate the relationship between IOP measurements and corneal parameters (central corneal thickness, corneal volume).
Main Methods:
- 118 healthy eyes underwent IOP measurement using both DCT and GAT.
- Central corneal thickness and corneal volume were measured using a Pentacam Scheimpflug camera.
- Measurements were taken in a specific sequence: Pentacam, DCT, then GAT after a 10-minute interval.
Main Results:
- DCT measurements (10.5–25.1 mm Hg) were significantly higher than GAT measurements (11–22 mm Hg).
- A statistically significant correlation was found between IOP measurement differences and central corneal thickness and corneal volume.
- No significant correlation was observed between IOP differences and corneal radius or age.
Conclusions:
- A discrete correlation exists between GAT and DCT, but DCT consistently yields higher IOP values.
- DCT may necessitate a re-evaluation of normal IOP ranges if considered a gold standard.
- The GAT and DCT devices are not interchangeable for IOP measurement.