Inter-observer variability of prostate delineation on cone beam computerised tomography images
E A White1, K K Brock, D A Jaffray
1The Radiation Medicine Programme, Princess Margaret Hospital, Toronto, Ontario, Canada.
Inter-observer variability in prostate gland contouring on cone beam CT images was assessed. While boundary definition showed variability, the impact on center of mass was minimal, supporting soft tissue guidance in radiotherapy.
Area of Science:
- Radiotherapy and Oncology
- Medical Imaging
- Radiation Oncology
Background:
- Image-guided radiotherapy (IGRT) relies on accurate target delineation.
- Cone beam computed tomography (CBCT) is increasingly used for soft tissue IGRT.
Purpose of the Study:
- To quantify inter-observer variability in prostate gland contouring on CBCT images.
- To assess the impact of this variability on radiotherapy planning and delivery.
Main Methods:
- Five genitourinary oncologists contoured prostate glands on five CBCT datasets.
- Variations in boundary delineation, isocenter placement, volume, and center of mass were measured.
- Finite element modeling was used to determine boundary definition variations.
Main Results:
- Average standard deviations for center of mass displacement were small (0.7-2.8 mm).
- Standard deviation for prostate volume determination was 8.93 cm³, with significant variability (3.98-19.00 cm³).
- Mean standard deviations for boundary displacements ranged from 1.8 mm to 3.6 mm.
Conclusions:
- Expert observers exhibited difficulty in defining the prostate peri-prostatic interface on CBCT.
- Inter-observer variability in prostate detection on CBCT is not prohibitive for soft tissue guidance.
- Rigorous training and protocol development are needed to mitigate potential systematic matching errors.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography