Inter-observer variability of prostate delineation on cone beam computerised tomography images
E A White1, K K Brock, D A Jaffray
1The Radiation Medicine Programme, Princess Margaret Hospital, Toronto, Ontario, Canada.
Aim:
To determine the inter-observer variability of defining the prostate gland on cone beam computerised tomography images for the purposes of image-guided radiotherapy.
Materials And Methods:
Five genitourinary oncologists contoured the prostate gland on five cone beam computerised tomography datasets. The variations in prostate boundary delineation and consequent isocentre placement between observers were measured. Variations in volume and centre of mass were calculated. The variation in boundary definition was determined with finite element modelling.
Results:
The average standard deviation for centre of mass displacements was small, measuring 0.7, 1.8 and 2.8mm in the left-right, anterior-posterior and superior-inferior directions, respectively. The standard deviation for volume determination was 8.93 cm(3) with large variability (3.98-19.00 cm(3)). The mean difference between the computerised tomography-derived volume and the mean cone beam-derived volume was 16% (range 0-23.7%). The mean standard deviations for left-right, anterior-posterior and superior-inferior boundary displacements were, respectively, 1.8, 2.1 and 3.6 mm. The maximum deviation seen was 9.7 mm in the superior direction.
Conclusion:
Expert observers had difficulty agreeing upon the location of the prostate peri-prostatic interface on the images provided. The effect on the centre of mass determination was small, and inter-observer variability for prostate detection on cone beam computerised tomography images is not prohibitive to the use of soft tissue guidance protocols. Potential exists for significant systematic matching errors, and points to the need for rigorous therapist image recognition training and development of guidance protocols before clinical implementation of soft tissue cone beam image guidance.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography