Real-time Pressure-volume Analysis of Acute Myocardial Infarction in Mice
Published on: July 2, 2018
Measuring left ventricular function in the normal, infarcted and CORM-3-preconditioned mouse heart using complex
James E Clark1, Anil Kottam, Roberto Motterlini
1Kings College London, Cardiovascular Division, London SE17EH, UK. james.2.clark@kcl.ac.uk
Journal of Pharmacological and Toxicological Methods
|December 9, 2008
Summary
A new complex admittance system (ADVantage system) accurately measures left ventricular (LV) function in mice, even in infarcted hearts. This method offers a reproducible and reliable alternative to traditional conductance catheters for assessing cardiac performance.
Area of Science:
- Cardiovascular Research
- Physiology
- Biomedical Engineering
Background:
- Miniaturized conductance catheters are used to assess left ventricular (LV) function in mice for cardiac interventions.
- Complex admittance offers real-time measurement correction, unlike traditional conductance catheters.
- Validating complex admittance for LV function assessment in normal and infarcted hearts is crucial.
Purpose of the Study:
- To validate the ADVantage system, a complex admittance-based method, for assessing LV function.
- To compare the ADVantage system with traditional conductance-derived pressure-volume (PV) systems.
- To evaluate the system's ability to detect LV dysfunction in a mouse model of myocardial infarction.
Main Methods:
- Direct comparison of the ADVantage system and a traditional PV system using a single conductance catheter in mice.
- Induction of focal myocardial ischemia-reperfusion injury to create dysfunctional LV.
- Measurement of cardiac function to differentiate normal from dysfunctional LV contractile performance.
Main Results:
- Complex admittance provides reproducible and reliable LV pressure-volume loops, comparable to conductance catheters.
- The ADVantage system recorded larger systolic LV cavity volumes than the traditional system.
- Significant LV dysfunction was observed post-coronary artery occlusion, ameliorated by CORM-3 (carbon monoxide-releasing molecule).
Conclusions:
- The ADVantage system effectively and reproducibly measures LV function and dysfunction in mice.
- This system eliminates the need for complex calibration interventions.
- The ADVantage system shows promise for accurate murine cardiac function assessment.