Related Experiment Video
Updated: Jun 26, 2026

08:57
Quantification of Optic Nerve Cross Sectional Area on MRI: A Novel Protocol using Fiji Software
Published on: September 4, 2021
Quantitative depth analysis of optic nerve head using stereo retinal fundus image pair.
Toshiaki Nakagawa1, Takayoshi Suzuki, Yoshinori Hayashi
1Gifu University, Graduate School of Medicine, 1-1 Yanagido, Gifu, Gifu 501-1194, Japan. tsnakagw@kowa.co.jp
Journal of Biomedical Optics
|January 7, 2009
Summary
This study presents an automatic method for measuring optic nerve head (ONH) depth from stereo fundus images, aiding glaucoma detection. The technique accurately quantifies ONH cup depth, showing good agreement with HRT measurements.
Area of Science:
- Ophthalmology
- Medical Imaging
- Computational Vision
Background:
- Glaucoma diagnosis relies on early detection of optic nerve head (ONH) changes.
- Accurate quantitative depth measurement of the ONH is crucial for assessing glaucoma progression.
- Existing methods may require specialized equipment or complex procedures.
Purpose of the Study:
- To develop and validate an automatic method for quantitative depth measurement of the optic nerve head (ONH) using stereo retinal fundus images.
- To assess the feasibility of using stereo fundus imaging for routine glaucoma screening and diagnosis.
- To compare the accuracy of the proposed depth measurement technique against a gold-standard imaging modality.
Main Methods:
- A five-step automatic reconstruction method was developed, including ONH region cutout, stereo image registration, disparity measurement, noise reduction, and depth calculation.
- Stereo retinal fundus images were acquired from 12 normal eyes.
- Depth measurements were performed using the proposed stereo imaging method and compared with those obtained from a Heidelberg Retina Tomograph (HRT).
Main Results:
- The proposed automatic method successfully reconstructed quantitative depth values of the ONH from stereo retinal fundus images.
- Depth measurements from the stereo fundus image pair showed high accordance with HRT measurements (r=0.80+/-0.15).
- The method demonstrated good performance in depth assessment of the ONH cup.
Conclusions:
- The developed automatic reconstruction method provides a reliable and quantitative measure of ONH depth from stereo retinal fundus images.
- This technique shows potential as an easy-to-handle tool for routine ophthalmological diagnosis and glaucoma screening.
- Stereo fundus imaging offers a promising, accessible approach for assessing ONH morphology relevant to glaucoma.