Detection of ventricular fibrillation with a ventricular monopolar catheter electrode
M B Voelz1, J L Wessale, L A Geddes
1William A. Hillenbrand Biomedical Engineering Center, Purdue University, West Lafayette, Indiana 47907.
Abstract:
This report describes the use of right ventricular impedance (RVZ) sensing with a monopolar electrode to identify the cessation of pumping with the onset of ventricular fibrillation. RVZ, ECG (lead II), and femoral artery pressure were monitored in seven anesthetized dogs. An impedance recorder (10-kHz, 100-microA peak-to-peak current) was used to measure RVZ between a monopolar, catheter-mounted electrode placed in the right ventricle and an indifferent electrode (8.5-cm diameter plate) sutured to the right chest wall. The catheter electrode was either 0.4 cm or 1.2 cm long and had a surface area of 0.50 cm2 or 1.50 cm2, respectively. A monopolar electrode was positioned at the apex of the right ventricle or midway between the apex and the tricuspid valve. The peak-to-peak amplitude of the pulsatile cardiac-induced impedance change was measured prior to and throughout an episode of ventricular fibrillation lasting up to 1 minute. On the average, the amplitude of the pulsatile RVZ signal after 10 seconds of fibrillation decreased by 79% of the prefibrillation amplitude. The 1.2-cm electrode located at the middle of the ventricle showed the largest reduction in pulsatile impedance, the amplitude being very small at 10 seconds. It is concluded that sensing RVZ by a monopolar electrode located on a catheter in the mid-ventricle can provide the mechanical information needed to identify the onset of ventricular fibrillation.