Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
High speed, wide velocity dynamic range Doppler optical coherence tomography (Part I): System design, signal
Abstract:
Improvements in real-time Doppler optical coherence tomography (DOCT), acquiring up to 32 frames per second at 250 x 512 pixels per image, are reported using signal processing techniques commonly employed in Doppler ultrasound imaging. The ability to measure a wide range of flow velocities, ranging from less than 20 microm/s to more than 10 cm/s, is demonstrated using an 1.3 microm DOCT system with flow phantoms in steady and pulsatile flow conditions. Based on full implementation of a coherent demodulator, four different modes of flow visualization are demonstrated: color Doppler, velocity variance, Doppler spectrum, and power Doppler. The performance of the former two, which are computationally suitable for real-time imaging, are analyzed in detail under various signal-to-noise and frame-rate conditions. The results serve as a guideline for choosing appropriate imaging parameters for detecting in vivo blood flow.