Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
Synchronized dual-polarity electrospray ionization mass spectrometry
Han-Kwang Chen1, Chia-Kai Chang, Chih-Che Wu
1Department of Chemistry, Chi Nan University, Nantou, Taiwan, Republic of China.
This study introduces dual-polarity electrospray ionization (DP-ESI), enabling simultaneous mass spectrometry analysis. This novel method broadens operational flow rates and enhances analytical capabilities for complex samples.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Ionization Techniques
Background:
- Electrospray ionization (ESI) is a crucial technique in mass spectrometry.
- Conventional ESI has limitations in flow rate range and sensitivity for certain applications.
- Dual-polarity (DP) ESI offers a potential advancement for ESI-MS.
Purpose of the Study:
- To describe the synchronized dual-polarity (DP) electrospray ionization (ESI) method.
- To demonstrate the first DP ESI mass spectra using two mass spectrometers.
- To evaluate the performance of DP ESI compared to conventional ESI.
Main Methods:
- Development of a single electrosprayer capable of producing stable double Taylor cones using counter electric voltages of opposite polarities.
- Optimization of extraction voltages and liquid flow rates for DP ESI.
- Acquisition of synchronized DP ESI mass spectra using electrospray calibrants in methanol solution.
Main Results:
- Stable double Taylor cones were successfully generated.
- DP ESI required higher extraction voltages but operated effectively over a wider liquid flow rate range (three times) than conventional ESI.
- Emission currents from the two cones neutralized in pure methanol, resulting in no net current drawn from the sprayer.
- Synchronized DP mass spectra were obtained, demonstrating the feasibility of the method.
- Conventional electrospray of bovine insulin showed satisfactory sensitivity and stability for routine analysis.
Conclusions:
- The synchronized DP-ESI method is a viable technique for simultaneous mass analysis.
- DP-ESI expands the operational parameters of ESI, particularly the liquid flow rate range.
- This method holds promise for enhanced analytical throughput and applications in mass spectrometry.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Tandem Mass Spectrometry
Mass Spectrum: Interpretation
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
MALDI-TOF Mass Spectrometry
Mass Spectrometers