Related Experiment Video
Updated: Jun 18, 2026

An Efficient and Rapid HPLC-QQQ-MS Method for the Quantitation of Tropane Alkaloids in Medicinal Plants
Published on: June 12, 2026
RP-HPLC method for simultaneous estimation of bisoprolol fumarate and hydrochlorothiazide in tablet formulation
Sneha J Joshi1, Pradnya A Karbhari, Suvarna I Bhoir
1C.B. Patel Research Centre, 3rd Floor, Bhaidas Hall, Vile Parle West, Mumbai 400056, India. sneha joshi@vsnl.net
Abstract:
A simple, precise and stability-indicating HPLC method was developed and validated for the simultaneous determination of bisoprolol fumarate and hydrochlorothiazide in pharmaceutical dosage form. The method involves the use of easily available inexpensive laboratory reagents. The separation was achieved on an Inertsil ODS 3V (25cmx4.6mm) 5microm column with isocratic flow. The mobile phase at a flow rate of 1.0mLmin(-1), consisted of 0.1M potassium dihydrogen phosphate buffer and acetonitrile (70:30, v/v). The UV detection was carried out at 228nm. A linear response was observed over the concentration range 2.5-50microgmL(-1) of bisoprolol fumarate and the concentration range 6.25-125microgmL(-1) of hydrochlorothiazide. Limit of detection and limit of quantitation for bisoprolol fumarate were 0.01 and 0.03microgmL(-1), respectively and for hydrochlorothiazide were 0.01 and 0.05microgmL(-1), respectively. The method was successfully validated in accordance to ICH guidelines acceptance criteria for specificity, linearity, accuracy, precision, robustness, ruggedness and system suitability. Individual drugs (bisoprolol fumarate and hydrochlorothiazide), their combinations and the tablets were exposed to thermal, photolytic, hydrolytic and oxidative stress conditions. The resultant stressed samples were analyzed by the proposed method. The method gave high resolution among the degradation products and the analytes. The peak purity of analyte peaks in the stressed samples was confirmed by photodiode array detector. The method was used for accelerated stability study on marketed and in-house formulations. The analysis concluded that the method was selective for simultaneous estimation of bisoprolol fumarate and hydrochlorothiazide and was stability-indicating.
More Related Videos
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018