[Comparative data regarding two HPLC methods for determination of isoniazid].

Daniela Gârbuleţ1, A F Spac, V Dorneanu

  • 1Facultatea de Farmacie, Disciplina Chimie analitică, Universitatea de Medicină si Farmacie "Gr.T. Popa" Iaşi.

Revista Medico-Chirurgicala a Societatii De Medici Si Naturalisti Din Iasi
|March 3, 2010
PubMed
Summary

Two validated High-Performance Liquid Chromatography (HPLC) methods were developed for determining isoniazid (isonicotinic acid hydrazide). Both methods demonstrated excellent precision, accuracy, and low detection limits for tablet analysis.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

IN VITRO MEASUREMENT OF TOTAL ANTIOXIDANT CAPACITY OF CRATAEGUS MACRACANTHA LODD LEAVES.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2016
Same author

Spectrophotometric method for estimation of bisoprolol fumarate in tablets.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2014
Same author

Evaluation of free radical scavenging activity of some antioxidants.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2014
Same author

Contributions to the phytochemical study of Bidens tripartitae herba from Romania. I. Tannins.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2014
Same author

Spectrometric determination of total antioxidant activity in chlorpromazine radical cation - ascorbic acid system.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2014
Same author

Spectrometric method for the assay of ramipril using molybdophosphoric acid.

Revista medico-chirurgicala a Societatii de Medici si Naturalisti din Iasi·2013

Area of Science:

  • Analytical Chemistry
  • Pharmaceutical Analysis

Background:

  • Isoniazid (isonicotinic acid hydrazide) is a crucial drug in tuberculosis treatment.
  • Accurate quantification of isoniazid in pharmaceutical formulations is essential for quality control.

Purpose of the Study:

  • To develop and validate two distinct High-Performance Liquid Chromatography (HPLC) methods for the determination of isoniazid.
  • To compare the performance of the two developed HPLC methods.

Main Methods:

  • Two HPLC methods were developed using a Waters 2695 liquid chromatograph with a UV detector.
  • Method I employed a Nucleosil 100-10 C18 column with an acetonitrile/oxalic acid mobile phase.
  • Method II utilized a Luna 100-5 C18 column with a methanol/acetate buffer mobile phase.

Related Experiment Videos

Last Updated: Jun 15, 2026

Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples
08:54

Validated LC-MS/MS Panel for Quantifying 11 Drug-Resistant TB Medications in Small Hair Samples

Published on: May 19, 2020

Main Results:

  • Both methods achieved high linearity with correlation coefficients of 0.9998 (I) and 0.9999 (II).
  • Excellent detection and quantitation limits were observed: 0.6 µg/mL (I) and 0.055 µg/mL (II) for detection; 1.9 µg/mL (I) and 0.2 µg/mL (II) for quantitation.
  • High accuracy (recoveries >101%) and precision (RSD < 2%) were confirmed for both methods.

Conclusions:

  • Both validated HPLC methods provide reliable and comparable results for isoniazid determination in tablets.
  • The developed methods are suitable for routine quality control analysis of isoniazid formulations.