Related Experiment Video
Updated: Jun 15, 2026

Low-cost Custom Fabrication and Mode-locked Operation of an All-normal-dispersion Femtosecond Fiber Laser for Multiphoton Microscopy
Published on: November 22, 2019
Construction of a femtosecond laser microsurgery system
Joseph D Steinmeyer1, Cody L Gilleland, Carlos Pardo-Martin
1Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, Massachusetts, USA.
Abstract:
Femtosecond laser microsurgery is a powerful method for studying cellular function, neural circuits, neuronal injury and neuronal regeneration because of its capability to selectively ablate sub-micron targets in vitro and in vivo with minimal damage to the surrounding tissue. Here, we present a step-by-step protocol for constructing a femtosecond laser microsurgery setup for use with a widely available compound fluorescence microscope. The protocol begins with the assembly and alignment of beam-conditioning optics at the output of a femtosecond laser. Then a dichroic mount is assembled and installed to direct the laser beam into the objective lens of a standard inverted microscope. Finally, the laser is focused on the image plane of the microscope to allow simultaneous surgery and fluorescence imaging. We illustrate the use of this setup by presenting axotomy in Caenorhabditis elegans as an example. This protocol can be completed in 2 d.