Simultaneous Laryngopharyngeal and Conventional Esophageal pH Monitoring
Published on: December 14, 2020
Wireless capsules for esophageal pH monitoring: are we placing them correctly?
Siva Doma1, Susmita Paladugu, Henry P Parkman
1Gastroenterology Section, Temple University School of Medicine, Philadelphia, PA 19140, USA.
Background:
Recommended Bravo capsule placement is 6 cm proximal to the squamocolumnar junction. This is because the junction resides 1 cm distal to the lower esophageal sphincter.
Aim:
To determine the positional accuracy of capsule placement compared with the ideal location in symptomatic patients.
Methods:
Retrospective analysis of consecutive symptomatic outpatients undergoing both capsule placement and esophageal manometry on the same day. Error in capsule placement (ECP) was calculated as the actual capsule position (ACP) minus the ideal capsule position (ICP) based on manometry results. An error in positioning < or =3 cm was deemed an accurate placement.
Results:
Accurate placement of the capsule occurred in 91/147 (62%) patients. In patients with inaccurate placement, 92.9% were placed >3 cm proximal to the ICP. Only longer esophageal length correlated with the severity of proximal misplacement. There was no relationship between severity of ECP and whether the study demonstrated acid reflux.
Conclusions:
Our results suggest that endoscopic misplacement of Bravo capsules is common. Capsule misplacement proximally occurred far more frequently than distal misplacement. Capsule misplacement was not associated with pH study results.
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Esophageal Achalasia
Esophageal Varices-II: Clinical Features and Management
In the initial assessment, a thorough review of the patient's medical history is vital to identify risk factors such as liver disease, alcohol abuse, or...
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...