Mebratu Madebo1, A Perkins, C Fox
1Department of Physical Sciences, Peter MacCallum Cancer Centre, Locked Bag 1 A'Beckett Street, Melbourne, VIC 8006, Australia. mebratu.madebo@petermac.org
Abstract:
Field junctions between megavoltage photon beams are important in modern radiotherapy for treatments such as head and neck and breast cancer. An electronic portal imaging device (EPID) may be used to study junction dose between two megavoltage X-ray fields. In this study, the junction dose was used to determine machine characteristics such as jaw positions and their reproducibility, collimator rotation and the effect of gantry rotation. All measurements were done on Varian linear accelerators with EPID (Varian, Palo Alto, CA). The results show reproducibility in jaw positions of approximately 0.3 mm for repeated jaw placement while EPID readings were reproducible within a standard deviation of 0.4% for fixed jaw positions. Junction dose also allowed collimator rotation error of 0.1 degrees to be observed. Dependence of junction dose on gantry rotation due to gravity was observed; the gravity effect being maximum at 180 degrees gantry angle (beam pointing up). EPIDs were found to be reliable tools for checking field junctions, which in turn may be used to check jaw reproducibility and collimator rotation of linacs.
More Related Videos
Related Concept Videos
X-ray Imaging
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Determination of Crystal Structures
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET