Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
An optimized online verification imaging procedure for external beam partial breast irradiation.
David J Willis1, Tomas Kron, Boon Chua
1Division of Radiation Oncology, Peter MacCallum Cancer Centre, East Melbourne, Victoria, Australia. David.Willis@petermac.org
Kilovoltage (kV) on-board imager (OBI) verification imaging using nonorthogonal kV image pairs is effective for partial breast irradiation. This method ensures accurate patient setup and clear visualization of surgical clips with minimal collision risk.
Area of Science:
- Radiation Oncology
- Medical Imaging
- Image-Guided Therapy
Background:
- External-beam partial breast irradiation requires precise patient setup.
- On-line verification imaging is crucial for accurate treatment delivery.
- Kilovoltage (kV) on-board imagers (OBI) offer potential for real-time imaging during radiotherapy.
Purpose of the Study:
- To evaluate the efficacy of a kV OBI-equipped linear accelerator for on-line verification imaging in partial breast irradiation.
- To optimize and assess imaging techniques for image quality, dose, and physical clearance.
- To determine the clinical applicability of nonorthogonal kV image pairs for online setup correction.
Main Methods:
- Optimization and assessment of kV imaging techniques using an anthropomorphic phantom for image quality and dose.
- Evaluation of physical clearance using a volunteer.
- Implementation of optimized nonorthogonal kV image pairs for online verification in 17 patients undergoing accelerated partial breast irradiation.
- Recording of imaging session duration, collision risk, visualization of surgical clips, and setup deviations.
Main Results:
- Nonorthogonal kV image pairs demonstrated optimal image quality, clearance, and dose.
- All 170 image sessions showed low collision risk.
- Surgical clips were well visualized in all cases where present (n=5).
- Average imaging session time was 6 minutes 16 seconds, with a learning curve effect.
- Online setup deviations up to 1.3 cm were detected and corrected.
Conclusions:
- Nonorthogonal kV image pairs provide effective and efficient online verification for partial breast irradiation.
- The technique allows for accurate online correction of patient setup deviations.
- Further multicenter studies are needed to assess generalizability and dependence on skilled staff.