Factors influencing pacemaker generator longevity: results from the complete automatic pacing threshold utilization
Lawrence S Rosenthal1, Steven Mester, Peter Rakovec
1UMassMemorial Medical Center, Worcester, Massachusetts, USA. rosenthl@ummhc.org
Pacing and Clinical Electrophysiology : PACE
|June 16, 2010
Summary
Automated atrial and right ventricular (RV) threshold algorithms in pacemakers are accurate and reliable. Features like threshold tracking and reduced pacing improve device longevity, with battery capacity being the key factor.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Modern pacemakers incorporate advanced features to enhance patient follow-up and optimize pacing parameters.
- Algorithms aim to reduce pacing output and percentages, thereby extending device longevity.
Purpose of the Study:
- To evaluate the accuracy and reliability of automated atrial and right ventricular (RV) threshold algorithms.
- To assess the projected longevity improvements associated with various pacemaker features and device characteristics.
Main Methods:
- Automated atrial and RV threshold measurements were compared against manual measurements at a 6-month follow-up.
- Projected device longevity was assessed with threshold-tracking features On versus Off.
- The impact of features reducing atrial and ventricular pacing, high impedance leads, and device size on longevity was investigated.
Main Results:
- Automated and manual measurements for atrial and RV thresholds demonstrated high concordance (98.8% and 98.7%, respectively).
- Thresholds remained stable, with most consecutive measurements within +/-0.25V.
- Threshold tracking, reduced ventricular pacing, and high impedance leads were associated with projected longevity increases of 0.8, 0.9, and 0.8 years, respectively. A 1-cc increase in device size projected a 2-year longevity improvement.
Conclusions:
- Automated atrial and RV algorithms are accurate and reliable for pacemaker threshold measurements.
- Threshold tracking, reduced ventricular pacing, and high impedance leads contribute to enhanced device longevity.
- Battery capacity emerged as the most significant determinant of projected pacemaker longevity.
Related Concept Videos
Pulse rhythm
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrophysiology of Normal Cardiac Rhythm
The normal cardiac rhythm is a synchronized electrical activity that facilitates the regular and coordinated contraction of the heart muscle. This process is essential for efficient blood circulation throughout the body. The fundamental elements involved in establishing and maintaining this rhythm include the unique electrical properties of cardiac muscle cells, the sinoatrial (SA) node's pacemaker function, the specialized conducting system, and the ionic mechanisms underlying each phase of...