Related Experiment Video
Updated: Dec 31, 2025

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Leak and obstruction with mask ventilation during simulated neonatal resuscitation
Kim Schilleman1, Ruben S Witlox, Enrico Lopriore
1Department of Pediatrics, Division of Neonatology, Leiden University Medical Center, Leiden, The Netherlands.
Objectives:
To evaluate mask technique during simulated neonatal resuscitation and test the effectiveness of training in optimal mask handling.
Study Design:
Seventy participants(consultants, registrars and nurses) from neonatal units were asked to administer positive pressure ventilation at a flow of 8 l/min and a frequency of 40-60/min to a modified leak free, term newborn manikin (lung compliance 0.5 ml/cm H(2)O) using a Neopuff T-piece device. Recordings were made (1) before training, (2) after training in mask handling and (3) 3 weeks later. Leak was calculated. Obstruction (tidal volume <60% of optimal tidal volume) and severe obstruction (<30% of optimal tidal volume) were calculated when leak was minimal.
Results:
For the 70 participants, median (IQR) leak was 71% (32-95%) before training, 10% (5-37%) directly after training and 15% (4-33%) 3 weeks later (p<0.001). When leak was minimal, gas flow obstruction was observed before, directly after training and 3 weeks later in 46%, 42% and 37% of inflations, respectively. Severe obstruction did not occur.
Conclusions:
Mask ventilation during simulated neonatal resuscitation was often hampered by large leaks at the face mask. Moderate airway obstruction occurred frequently when effort was taken to minimise leak. Training in mask ventilation reduced mask leak but should also focus on preventing airway obstruction.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: