Related Experiment Video
Updated: Jun 10, 2026

Amplification of Escherichia coli in a Continuous-Flow-PCR Microfluidic Chip and Its Detection with a Capillary Electrophoresis System
Published on: November 21, 2023
A microfluidic device integrated with multichamber polymerase chain reaction and multichannel separation for genetic
Xiaoyan Pan1, Lei Jiang, Kaiying Liu
1Dalian Institute of Chemical Physics, Chinese Academy of Sciences, 457 Zhongshan Road, Dalian 116023, PR China.
Abstract:
This work describes a microfluidic device integrated with multichamber polymerase chain reaction (PCR) and multichannel separation for parallel genetic analysis. The microdevice consists of three functional units: temperature control, multiple PCR (four chambers PCR), and multiple channel separation (four separation channels, each channel connected to a PCR chamber). Platinum (Pt)/titanium (Ti) microheater was used to ensure homogeneous temperature field, and Pt-chip sensor was used for temperature monitoring. The interface between chip-PCR and chip separation was simplified by connecting the PCR chamber with separation channel directly. After chip-PCR, PCR products were introduced into parallel separation channels for subsequent separation/detection by applying an electric field automatically. This microdevice was successfully applied for detection of pathogens including hepatitis B virus (HBV) and Mycobacterium tuberculosis (MTB), and genotyping of human leucocyte antigen (HLA)-B27 as well, demonstrating the feasibility of the integrated microdevice for parallel genetic analysis.