Radiation dose for routine clinical adult brain CT: Variability on different scanners at one institution
Tracy A Jaffe1, Jenny K Hoang, Terry T Yoshizumi
1Department of Radiology, Duke University Medical Center, Durham, NC 27710, USA. jaffe002@mc.duke.edu
Objective:
The purpose of this study was to determine, using an anthropomorphic phantom, whether patients are subject to variable radiation doses based on scanner assignment for routine CT of the brain.
Materials And Methods:
Twenty metal oxide semiconductor field effect transistor dosimeters were placed in the brain of a male anthropomorphic phantom scanned three times with a routine clinical brain CT protocol on four scanners from one manufacturer in four configurations and on one 64-MDCT scanner from another manufacturer. Absorbed organ doses were measured for skin, cranium, brain, lens of the eye, mandible, and thyroid. Effective dose was calculated on the basis of the dose-length product recorded on each scanner.
Results:
Organ dose ranges were as follows: cranium, 2.57-3.47 cGy; brain, 2.34-3.78 cGy; lens, 2.51-5.03 cGy; mandible 0.17-0.48 cGy; and thyroid, 0.03-0.28 cGy. Statistically significant differences between scanners with respect to dose were recorded for brain and lens (p < 0.05). Absorbed doses were lowest on the single-detector scanner. In the comparison of MDCT scanners, the highest doses were found on the 4-MDCT scanner and the dual-source 64-MDCT scanner not capable of gantry tilt. Effective dose ranged from 1.22 to 1.86 mSv.
Conclusion:
According to the phantom data, patients are subject to different organ doses in the lens and brain depending on scanner assignment. At our institution with existing protocols, absorbed doses at brain CT are lowest with the single-detector CT scanner, followed by MDCT scanners capable of gantry tilt. On scanners without gantry tilt, CT of the brain should be performed with careful head positioning and shielding of the orbits. These precautions are especially true for patients who need repeated scanning and for pediatric patients.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System V: CT