Related Experiment Video
Updated: Jun 10, 2026

Monitoring Protein Adsorption with Solid-state Nanopores
Published on: December 2, 2011
Thick Silicon Double-Sided Strip Detectors for Low-Energy Small-Animal SPECT
Sepideh Shokouhi1, Benjamin S McDonald, Heather L Durko
1Institute of Imaging Science, Vanderbilt University, Nashville, TN 37232-2310 USA.
Abstract:
This work presents characterization studies of thick silicon double-sided strip detectors for a high-resolution small-animal SPECT. The dimension of these detectors is 60.4 mm × 60.4 mm × 1 mm. There are 1024 strips on each side that give the coordinates of the photon interaction, with each strip processed by a separate ASIC channel. Our measurement shows that intrinsic spatial resolution equivalent to the 59 microm strip pitch is attainable. Good trigger uniformity can be achieved by proper setting of a 4-bit DAC in each ASIC channel to remove trigger threshold variations. This is particularly important for triggering at low energies. The thick silicon DSSD (Double-sided strip detector) shows high potential for small-animal SPECT.