Isocentric rotational performance of the Elekta Precise Table studied using a USB-microscope.
Hans L Riis1, Sune J Zimmermann, Poul Riis
1Radiofysisk Laboratorium, Odense University Hospital, Sdr. Boulevard 29, DK-5000 Odense C, Denmark. hans.lynggaard.riis@ouh.regionsyddanmark.dk
Physics in Medicine and Biology
|November 25, 2010
Summary
This study evaluated the Elekta Precise Table
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Accurate patient positioning is critical in radiation therapy.
- The performance of treatment tables affects dose delivery precision.
Purpose of the Study:
- To investigate the isocentric three-dimensional performance of the Elekta Precise Table.
- To develop and validate a new elastic model for the treatment table.
- To propose a method for optimal table positioning relative to the accelerator.
Main Methods:
- A pointer attached to the radiation head and a USB-microscope on the treatment tabletop were used for precise measurements.
- Table performance was mapped under various angles and load conditions.
- A new elastic model was developed and validated using six geometric and elastic parameters calculated via linear least squares fitting.
Main Results:
- The USB-microscope achieved a measurement resolution of 0.01 mm.
- The developed elastic model accurately described table performance, with minor deviations attributed to backlash.
- A novel method for optimal table positioning was presented.
Conclusions:
- The study provides a detailed performance analysis of the Elekta Precise Table.
- Incorporating geometric and elastic models into the dose planning system is recommended to eliminate systematic errors and enhance positioning accuracy.