Related Experiment Video
Updated: Jun 3, 2026

Fentanyl Analog Screening using LC-TIMS-TOF MS/MS
Published on: November 8, 2024
Simultaneous determination of alfentanil and midazolam in human plasma using liquid chromatography and tandem mass
Thomas Kim1, Amy London, Evan D Kharasch
1Division of Clinical and Translational Research, Department of Anesthesiology, Washington University in St. Louis, St. Louis, MO 63110, United States.
Abstract:
A fast, sensitive and selective liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the determination of alfentanil and midazolam in human plasma has been developed and validated. Alfentanil and midazolam were extracted from plasma using a mixed-mode cation exchange solid phase extraction method, with recoveries of both compounds greater than 80% at 3 different concentrations (1, 10 and 100ng/ml). Compounds were analyzed on a C(18) column with a water and methanol mobile phase gradient with acetic acid as an additive, at a flow rate of 0.3ml/min. The working assay range was linear from 0.25 to 100ng/ml for each compound. The signal to noise ratio was 80 and 40 for alfentanil and midazolam, respectively, at the lowest concentration calibration standard, with less than 10% matrix suppression by human plasma at this concentration. Alfentanil and midazolam were stable in human plasma during storage at -80°C, processing, and analysis. The procedure was validated and applied to the analysis of plasma samples from healthy human subjects administered oral and intravenous alfentanil and midazolam.
Related Concept Videos
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Types of Detectors
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...