Related Experiment Video
Updated: May 30, 2026

3D Cine Magnetic Resonance Imaging of Respiratory Motion in Mechanically Ventilated Mice and Rats
Published on: September 19, 2025
"MR conditional" respiratory ventilator system incident in a 3-T MRI environment
Kathryn L Z Greenberg1, Jeffrey Weinreb, Frank G Shellock
1Department of Radiology, Yale New Haven Hospital, New Haven, CT 06510, USA.
Abstract:
The misunderstanding of the labeling for an "MR conditional" respiratory ventilator system resulted in the placement of this device, which had substantial ferromagnetic components, too close to a 3-T magnetic resonance (MR) system, causing a projectile incident. Magnetic resonance imaging health care professionals should be aware of the potentially dangerous consequences when such medical devices that have Food and Drug Administration-approved MR conditional labeling are brought into the MR system room. Recommendations to prevent future incidents are provided herein.
Related Concept Videos
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System IV: CMRI
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Acute Respiratory Failure-III
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings