A Guide to Build a Highly Inclined Swept Tile Microscope for Extended Field-of-view Single-molecule Imaging
Published on: April 8, 2019
Time-varying view angle tilting with spiral readout gradients
Sang-Young Zho1, Dong-Hyun Kim
1Department of Electrical and Electronic Engineering, Yonsei University, Seoul, Korea.
This study explores using spiral trajectories in spin-echo imaging to speed up 3D scans. Results show spiral readouts maintain image quality comparable to traditional methods when acquisition times are similar.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Biomedical Engineering
- Medical Physics
Background:
- Conventional spin-echo sequences with view angle tilting correct in-plane distortions from chemical shift and B0 inhomogeneity.
- Extending these sequences to 3D applications necessitates efficient k-space sampling to reduce scan times.
- Spiral trajectories offer high k-space acquisition efficiency, making them a potential alternative to Cartesian sampling.
Purpose of the Study:
- To investigate the feasibility and performance of spiral trajectories within a spin-echo pulse sequence incorporating a view angle tilting gradient.
- To assess the impact of spiral readout gradients on image quality in the presence of view angle tilting.
- To determine if spiral trajectories can effectively reduce scan time in 3D MRI while maintaining diagnostic image quality.
Main Methods:
- Development and implementation of a spin-echo pulse sequence utilizing spiral readout gradients combined with a view angle tilting gradient.
- Conducting computer simulations to model the behavior of spiral trajectories under these conditions.
- Performing phantom and in vivo experiments to validate the simulation results and assess practical performance.
Main Results:
- Computer simulations and experimental data confirmed the successful integration of spiral readout gradients with view angle tilting.
- Image quality was found to be comparable between spiral and Cartesian readout methods when readout times were similar.
- The study demonstrated the potential for spiral trajectories to accelerate 3D MRI acquisition within this framework.
Conclusions:
- Spiral trajectories are applicable to spin-echo sequences with view angle tilting for 3D MRI.
- This approach offers a viable method for reducing scan times without compromising image quality, provided readout duration is maintained.
- The findings support the use of efficient k-space sampling strategies in advanced MRI techniques.
More Related Videos
06:30Assessment of Static Graviceptive Perception in the Roll-Plane using the Subjective Visual Vertical Paradigm
Published on: April 28, 2020
09:01Gain-compensation Methodology for a Sinusoidal Scan of a Galvanometer Mirror in Proportional-Integral-Differential Control Using Pre-emphasis Techniques
Published on: April 4, 2017
Related Concept Videos
Gyroscope: Precession
Galvanometer
The galvanometer consists of two concave-shaped permanent magnets, providing a uniform radial magnetic field in the annular region. In the center, a pivoted coil of fine copper wire is placed in the uniform magnetic...
Rotation with Constant Angular Acceleration - II
The first...
Angle of Twist: Problem Solving
Rotation with Constant Angular Acceleration - I
Using our intuition, we can begin to see how rotational quantities such as angular displacement, angular velocity, angular acceleration, and time are related to one another. For example, if a flywheel...
Gyroscope