Related Experiment Video
Updated: May 25, 2026

Tachycardia-Induced Cardiomyopathy As a Chronic Heart Failure Model in Swine
Published on: February 17, 2018
Atrial lead malfunction presenting as new onset pacemaker-mediated tachycardia
S Serge Barold1, Roland X Stroobandt
1The Florida Heart Rhythm Institute, Tampa, FL, USA. ssbarold@aol.com
Abstract:
This report describes the de novo occurrence of pacemaker-mediated tachycardia (PMT) in a patient with a dual-chamber implantable cardioverter-defibrillator and stable retrograde ventriculoatrial conduction time. The same rate-adaptive post-ventricular atrial refractory period (PVARP) duration had previously prevented PMT. Oversensing of atrial false signals from a defective lead shortened the PVARP with consequent sensing of retrograde conduction.