Related Experiment Video
Updated: May 25, 2026

Neuronavigation and Laparoscopy Guided Ventriculoperitoneal Shunt Insertion for the Treatment of Hydrocephalus
Published on: October 14, 2022
Micro-fabricated shunt to mimic arachnoid granulations for the treatment of communicating hydrocephalus
Francis Kralick1, Jonghyun Oh, Tim Medina
1Department of Neurological Surgery, Drexel University College of Medicine, Philadelphia, PA 19107, USA. fkralick@yahoo.com
Abstract:
Hydrocephalus is the abnormal accumulation of cerebrospinal fluid (CSF) within the confines of the skull that if left untreated results in significant morbidity and mortality. The treatment for hydrocephalus has remained essentially unchanged for over 50 years. It was a technological advance in materials that allowed John Holter, in conjunction with neurosurgeons Spitzer and Nulsen, to devise a valve and shunt system that diverted excess CSF from the ventricular space to the peritoneum. This ventriculo-peritoneal (VP) shunt is far from ideal, with problems associated with under/over shunting, mechanical mismatch, infection, high failure rates, disconnection and erosion. With the advances in the field of micro-fabrication and micro-machines we propose an innovative shunt system that would mimic the function of arachnoid granulations. This micro-fabricated shunting device, or micro-mechanical arachnoid granulation (MAG), consists of a multiplicity of micro-valves each 210 μm in diameter that each adhere to individual micro-needles. This work demonstrates the design and initial test results of the micro-valve with parameters for low cracking pressure, optimal flow rate, and reflux that would mimic the function of the native arachnoid granulations.