Clinical comparison of ICD detection algorithms that include rapid-VT zones.
David S Hirsh1, Larry A Chinitz, Neil E Bernstein
1Leon H. Charney Heart Rhythm Center, New York University School of Medicine, New York, New York, USA.
The Boston Scientific implantable cardioverter defibrillator (ICD) algorithm shows a longer time to detect ventricular fibrillation (VF), delaying device charging. This may impact the safety of antitachycardia pacing (ATP) for treating rapid ventricular tachycardia (VT).
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Antitachycardia pacing (ATP) for ventricular tachycardia (VT) relies on implantable cardioverter defibrillators (ICDs) distinguishing VT from ventricular fibrillation (VF).
- Accurate and rapid VT/VF discrimination is crucial for timely and appropriate device therapy.
Purpose of the Study:
- To compare the time to device charging following VF induction across different ICD multizone detection algorithms.
- To evaluate the performance of VT/VF discrimination algorithms from Boston Scientific, Medtronic, and St. Jude Medical devices.
Main Methods:
- A prospective cohort study of 62 patients undergoing ICD implantation.
- Induction of VF at the time of implantation to test device algorithms.
- Analysis of time to device charging and therapy initiation for Boston Scientific, Medtronic, and St. Jude Medical devices.
Main Results:
- Mean time to charging was significantly prolonged for the Boston Scientific group (4.24 seconds) compared to Medtronic (3.99 seconds) and St. Jude Medical (3.00 seconds).
- A higher percentage of tests showed charge initiation >5 seconds from VF induction in the Boston Scientific group (19%) versus Medtronic (4%) and St. Jude Medical (0%).
- Antitachycardia pacing (ATP) was initiated first in 9.4% of tests with the Boston Scientific device.
Conclusions:
- The Boston Scientific multizone VT/VF discrimination algorithm leads to a prolonged time for VF detection and subsequent device charging.
- Further research is necessary to determine the clinical significance of these prolonged detection times.
Related Concept Videos
Automated Microbial Diagnostics
Receiver Operating Characteristic Plot
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Sensitivity, Specificity, and Predicted Value
Sensitivity is the...