Related Experiment Video
Updated: May 23, 2026

Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Influence of metal artifacts on in vivo micro-CT for orthodontic mini-implants
Ken-ichiro Ejima1, Saori Omasa, Mitsuru Motoyoshi
1Department of Oral and Maxillofacial Radiology, Nihon University, School of Dentistry, Tokyo, Japan. ejima@dent.nihon-u.ac.jp
Abstract:
This study aimed to show the effects of metal artifacts on the in vivo micro-CT of mini-implants by measuring bone volume. We drilled a hole in the cortical bone of a rat tibia and embedded a titanium orthodontic mini-implant (diameter, 1.5 mm) in the hole. Twelve individually weighed hydroxyapatite grains (HA grains) were placed around the implant either by one dentist (method 1) or separately by 12 dentists (method 2). In vivo micro-CT was used to scan the model after placement of each grain to measure increases and decreases in bone volume voxel number. The subtracted bone voxel volume increased with HA weight in both methods. Simple linear regression analysis showed a significant correlation between weight and volume in both methods (method 1: regression coefficient: 516.502, P < 0.05; method 2: regression coefficient: 4837.432, P < 0.05). Metal artifacts did not appear to influence measurements of bone volume, although further studies are required to determine the effect of thicker implants.

