Measurement of the 17O-excess (Δ17O) of tropospheric ozone using a nitrite-coated filter
Related Experiment Video
Updated: May 23, 2026

Automated, High-resolution Mobile Collection System for the Nitrogen Isotopic Analysis of NOx
Published on: December 20, 2016
William C Vicars1, S K Bhattacharya, Joseph Erbland
1Laboratoire de Glaciologie et Géophysique de l'Environnement, Université Joseph Fourier-Grenoble 1/CNRS, Grenoble, France. william.vicars@lgge.obs.ujf-grenoble.fr
Rationale:
The (17)O-excess (Δ(17)O) of tropospheric ozone (O(3)) serves as a useful marker in studies of atmospheric oxidation pathways; however, due to the complexity and expense of currently available analytical techniques, no systematic sampling campaign has yet been undertaken and natural variations in Δ(17)O(O(3)) are therefore not well constrained.
Methods:
The nitrite-coated filter method is a new technique for O(3) isotope analysis that employs the aqueous phase NO(2)(-) + O(3) → NO(3)(-) + O(2) reaction to obtain quantitative information on O(3) via the oxygen atom transfer to nitrate (NO(3)(-)). The triple-oxygen isotope analysis of the NO(3)(-) produced during this reaction, achieved in this study using the bacterial denitrifier method followed by isotope-ratio mass spectrometry (IRMS), directly yields the Δ(17)O value transferred from O(3). This isotope transfer process was investigated in a series of vacuum-line experiments, which were conducted by exposing coated filters to O(3) of various known Δ(17)O values and then determining the isotopic composition of the NO(3)(-) produced on the filter.
Results:
The isotope transfer experiments revealed a strong linear correlation between the Δ(17)O of the O(3) produced and that of the oxygen atom transferred to NO(3)(-), with a slope of 1.55 for samples with bulk Δ(17)O(O(3)) values in the atmospheric range (20-40‰). This finding is in agreement with theoretical postulates that place the (17) O-excess on only the terminal oxygen atoms of ozone. Ambient measurements yield average Δ(17)O(O(3))(bulk) values in agreement with previous studies (22.9 ± 1.9‰).
Conclusions:
The nitrite-coated filter technique is a sufficiently robust, field-deployable method for the determination of the triple-oxygen isotopic composition of tropospheric O(3). Further ambient measurements will undoubtedly lead to an improved quantitative view of natural Δ(17)O(O(3)) variation and transfer in the atmosphere.