Related Experiment Video
Updated: May 21, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
The 37-year durability of a Björk-Shiley Delrin-disc aortic valve prosthesis
Fabrizio Sansone1, Edoardo Zingarelli, Guglielmo Mario Actis Dato
1Division of Cardiac Surgery, Mauriziano Hospital Umberto I, L.go Turati 62, 10135 Turin, Italy. fabrisans@katamail.com
Abstract:
We report the exceptional longevity of a Björk-Shiley Delrin-disc prosthetic aortic valve that had been implanted in a man who underwent surgical correction of an ascending aortic aneurysm 37 years later. Upon explantation of the valve, the Delrin disc had only shallow abrasion on the ventricular surface, and none on the aortic surface. We discuss the soundness and durability of this valve in our patient, in contrast with its short functional prosthetic life in other patients. The 37-year lifespan of this patient's Björk-Shiley Delrin-disc valve is among the longest reported.