[Determination of hydrogen selenide in workplace air with atomic fluorescence]
Wen-hua Qin1, Yan-yan Lu, Qi Geng
1Henan Institute of Occupational Medicine, Zhengzhou, 450052, China.
Objective:
To determined hydrogen selenide in workplace air with atomic fluorescence.
Method:
Hydrogen selenide were sampled with 0.1 mol/L sodium hydroxide solution in multi-hole absorbing tubes. The sampled absorbing solution were digested with (9+1) nitric acid/perchloric acid. The selenide in sample were reduced by potassium borohydride in 5.0% hydrochloride solution and determined with atomic fluorescence.
Results:
There was a good linearity (r=0.9999) over the concentration of 0-150 microg/L, The precision of low, middle and high concentration were 3.1%, 7.4% and 6.7%, respectively. The sample collection rate can reach 99%.
Conclusion:
The method was accurate and sensitive to detect hydrogen selenide in workplace air.