X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Radiation dose and physical image quality in 128-section dual-source computed tomographic coronary angiography: a
Kosuke Matsubara1, Haruka Koshida, Keita Sakuta
1Department of Quantum Medical Technology, Faculty of Health Sciences, Kanazawa University, Kanazawa, Japan. matsuk@mhs.mp.kanazawa-u.ac.jp
Abstract:
One-hundred-and-twenty-eight-section dual X-ray source computed tomography (CT) systems have been introduced into clinical practice and have been shown to increase temporal resolution. Higher temporal resolution allows low-dose spiral mode at a high pitch factor during CT coronary angiography. We evaluated radiation dose and physical image qualities in CT coronary angiography by applying high-pitch spiral, step-and-shoot, and low-pitch spiral modes to determine the optimal acquisition mode for clinical situations. An anthropomorphic phantom, small dosimeters, a calibration phantom, and a microdisc phantom were used to evaluate the radiation doses absorbed by thoracic organs, noise power spectrums, in-plane and z-axis modulation transfer functions, slice sensitivity profiles, and number of artifacts for the three acquisition modes. The high-pitch spiral mode had the advantage of a small absorbed radiation dose, but provided low image quality. The low-pitch spiral mode resulted in a high absorbed radiation dose of approximately 200 mGy for the heart. Although the absorbed radiation dose was lower in the step-and-shoot mode than in the low-pitch spiral mode, the noise power spectrum was inferior. The quality of the in-plane modulation transfer function differed, depending on spatial frequency. Therefore, the step-and-shoot mode should be applied initially because of its low absorbed radiation dose and superior image quality.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...