Related Experiment Video
Updated: May 14, 2026

06:55
Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Technique for measuring forward light scatter in intraocular lenses
Kamal K Das1, John C Stover, Jim Schwiegerling
1Alcon Laboratories, Inc., Fort Worth, Texas 76134-2099, USA. kamal.das@alconlabs.com
Journal of Cataract and Refractive Surgery
|February 12, 2013
Summary
A new technique quantifies light scattering in intraocular lenses (IOLs). Artificially aged IOLs showed increased scatter due to nanoglistenings, but vision impact appears minimal.
Area of Science:
- Ophthalmology and optical engineering.
- Materials science of intraocular lenses.
Background:
- Intraocular lenses (IOLs) are crucial for vision restoration after cataract surgery.
- Understanding light scattering in IOLs is essential for assessing optical performance and longevity.
- Previous methods may not fully capture forward light scattering characteristics.
Purpose of the Study:
- To develop and validate a novel technique for measuring forward light scattering in intraocular lenses (IOLs).
- To quantify light scatter in both new and artificially aged IOLs.
- To investigate the relationship between IOL aging, nanoglistenings, and light scatter patterns.
Main Methods:
- Utilized a scatterometer with laser sources (488 nm and 633 nm) to measure forward-scattered light from IOLs.
- Employed a goniometer setup with a detector to capture scatter distribution profiles.
- Examined both new and artificially aged IOLs immersed in balanced salt solution.
Main Results:
- Artificially aged IOLs exhibited increased forward light scatter compared to new IOLs.
- Subsurface nanoglistenings, consistent with Rayleigh-type scattering, were observed in aged IOLs.
- The measured increase in scatter from nanoglistenings was not substantial enough to likely impair vision.
Conclusions:
- A reliable technique for quantifying forward and back light scatter from IOLs has been successfully developed.
- The study demonstrates that artificial aging can induce nanoglistenings and alter light scattering properties of IOLs.
- The developed technique provides a valuable tool for IOL quality control and long-term performance assessment.