Interaction between cardioverter defibrillator and enhanced external counterpulsation device
Omer Celik1, Alper Aydin, Mustafa S Yilmazer
1Department of Cardiology, Istanbul Mehmet Akif Ersoy, Thoracic and Cardiovascular Surgery, Training and Research Hospital, Halkali, Istanbul, Turkey.
External enhanced counterpulsation (EECP) is safe for patients with implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy with defibrillator (CRT-D) devices. However, some CRT-D patients may experience QRS sensing issues or inappropriate sinus tachycardia during EECP.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Implanted cardiac devices like ICDs and CRT-Ds require careful consideration regarding electromagnetic interference from other medical equipment.
- External enhanced counterpulsation (EECP) is a non-invasive circulatory support therapy.
Purpose of the Study:
- To evaluate potential device interactions between implantable cardioverter defibrillators (ICDs) and external enhanced counterpulsation (EECP).
Main Methods:
- Twenty-one patients with ICD or CRT-D were monitored during two 5-minute EECP sessions.
- Device interrogations were performed before, during, and after EECP to assess for interactions.
Main Results:
- No inappropriate sensing or noise was detected during EECP sessions.
- No significant changes in electrode impedance, pacing, or sensing values were observed.
- Significant heart rate differences occurred between rate-responsive modes (P=0.001).
- Four CRT-D patients with unipolar pacing experienced QRS sensing problems, hindering counterpulsation.
Conclusions:
- EECP is generally safe for patients with ICDs and CRT-Ds.
- Caution is advised for CRT-D patients with the rate-responsive mode on due to potential sinus tachycardia.
- Unipolar sensing modes in CRT-D patients may lead to QRS sensing issues, impacting EECP synchronization and efficacy.
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiopulmonary Resuscitation IV: Pharmacological Management