Application of Optical Coherence Tomography to a Mouse Model of Retinopathy
Published on: January 12, 2022
Retinal thickness and volume measured with enhanced depth imaging optical coherence tomography
Sung Yong Park1, Sung Min Kim1, Yun-Mi Song2
1Department of Ophthalmology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, South Korea.
American Journal of Ophthalmology
|June 18, 2013
Summary
Enhanced depth imaging (EDI) optical coherence tomography (OCT) shows high agreement with conventional OCT for measuring retinal thickness and volume. This EDI OCT raster scan method is reliable for evaluating the retina and choroid in healthy eyes and those with chorioretinal disorders.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging
Background:
- Spectral-domain optical coherence tomography (OCT) is crucial for retinal imaging.
- Enhanced depth imaging (EDI) is a newer OCT protocol.
- Comparing EDI OCT with conventional OCT is important for clinical application.
Purpose of the Study:
- To compare retinal thickness and volume measurements obtained via EDI OCT and conventional OCT.
- To assess the agreement between these two OCT methods.
Main Methods:
- Retrospective analysis of 20 healthy subjects and 35 patients with chorioretinal diseases.
- Spectral-domain OCT performed using both conventional and EDI raster scan protocols.
- Measurement of retinal thickness and volume in 9 Early Treatment Diabetic Retinopathy Study subfields; statistical analysis using intraclass correlation coefficients and Bland-Altman plots.
Main Results:
- High intraclass correlation coefficients (0.861–0.995) between conventional and EDI OCT for retinal thickness and volume.
- Narrow 95% limits of agreement for central subfield measurements (-14.52 to 12.88 μm for thickness, -0.014 to 0.013 mm³ for volume).
- Statistically insignificant differences in segmentation error rates between protocols, except in cases of reticular pseudodrusen.
Conclusions:
- EDI OCT raster scan demonstrates high agreement with conventional OCT for retinal and choroidal measurements.
- EDI OCT is a reliable method for evaluating both retina and choroid in healthy individuals and patients with certain chorioretinal conditions.
- The findings support the clinical utility of EDI OCT in ophthalmology.