Estimation of sertraline by chromatographic (HPLC-UV273 nm) technique under hydrolytic stress conditions
Related Experiment Video
Updated: May 10, 2026

Thermostabilization, Expression, Purification, and Crystallization of the Human Serotonin Transporter Bound to S-citalopram
Published on: November 27, 2016
Md Akhlaquer Rahman1, Zeenat Iqbal, Mohd Aamir Mirza
1Faculty of Pharmacy, Integral University, Kursi Road, Lucknow - 226 026, India.
Purpose:
In this paper, simple, specific and accurate RP-HPLC method was developed in order to study decomposition of sertraline (SRT) under the hydrolytic stress conditions (acid, neutral, alkaline and oxidative).
Materials And Methods:
The best separation of SRT and its degradation products were achieved on reverse phase LiChoCART with Purospher (RP-18e) column. The mobile phase was composed of methanol/water (75:25, v/v). The detection wavelength was 273 nm. The method was validated and response was found to be linear in the drug concentration range of 10-200 μg ml-1 with correlation coefficient of 0.998.
Results:
The RSD values for intra- and inter-day precision were < 0.65 and < 0.72%, respectively. Employing RP-HPLC method, degradation products were detected in the exposed samples.
Conclusion:
IT WAS FOUND THAT THE SUSCEPTIBILITY OF SRT TO HYDROLYTIC DECOMPOSITION INCREASED IN FOLLOWING MANNER: Neutral condition < alkaline condition < acid condition < oxidative condition.