Electron-impact ionization detection of scopolamine by gas chromatography-mass spectrometry in rat plasma and brain
Related Experiment Videos
J Deutsch1, T T Soncrant, N H Greig
1Laboratory of Neurosciences, National Institute on Aging, National Institutes of Health, Bethesda, MD 20892.
Abstract:
Scopolamine, a muscarinic receptor antagonist, was measured in rat plasma and brain using gas chromatography-mass spectrometry with electron-impact ionization detection. Extracted scopolamine was either directly derivatized or first hydrolyzed to scopine, then derivatized with heptafluorobutyric anhydride, separated on a capillary column and quantified by mass fragmentography. Electron-impact ionization produced a common fragment peak at m/z 138 that was monitored along with trideuterated scopolamine, an internal standard (m/z 141). The method can be used to measure scopolamine concentrations of 2 ng/ml in rat plasma and 20 ng/g in rat brain.