Related Experiment Video
Updated: May 8, 2026

A Novel Approach for the Administration of Medications and Fluids in Emergency Scenarios and Settings
Published on: November 9, 2016
Transonic, thermodilution, or ionic dialysance to manage vascular access: which method is best?
Bouchra Badr1, Pierre Bories, Richard Marais
1Department of Nephrology, Dialysis and Organ Transplantation, CHU Rangueil, Toulouse, France.
Monitoring vascular access blood flow (Qa) with integrated devices like thermodilution and ionic dialysance is effective. Thermodilution offers better precision and correlation compared to saline dilution for predicting dysfunction and preventing thrombosis.
Area of Science:
- Nephrology
- Vascular Surgery
- Biomedical Engineering
Background:
- Regular monitoring of vascular access blood flow (Qa) is crucial for predicting dysfunction and preventing thromboses.
- Integrated access flow devices offer potential for easier and more cost-effective monitoring.
Purpose of the Study:
- To compare the accuracy and reproducibility of two integrated access flow devices: thermodilution (Qa-BTM) and ionic dialysance (Qa-ID).
- To evaluate these devices against the gold standard saline dilution method (Qa-T) for measuring Qa in hemodialysis patients.
Main Methods:
- Sequential measurements of access flow using Qa-T, Qa-BTM, and Qa-ID were performed in 24 long-term hemodialysis patients.
- Statistical analyses included Bland-Altman, linear regression (r²), and intraclass correlation coefficients (ICC) to assess reproducibility, correlations, and concordance.
- Recirculation was measured using both saline dilution and thermodilution methods.
Main Results:
- Average access flow measurements were comparable across all three methods (Qa-T: 1549 mL/min, Qa-BTM: 1530 mL/min, Qa-ID: 1619 mL/min).
- Thermodilution (Qa-BTM) demonstrated superior precision and concordance with saline dilution (ICC=0.99, r²=0.98, bias=-19 mL/min) compared to ionic dialysance (ICC=0.75, r²=0.65, bias=+69 mL/min).
- Reproducibility of thermodilution and ionic dialysance was not significantly different from saline dilution; recirculation measurements were identical (0%).
Conclusions:
- Integrated access flow measurement devices, specifically thermodilution, are reliable alternatives to saline dilution for monitoring vascular access.
- Thermodilution offers enhanced precision and correlation, simplifying vascular access monitoring without additional costs or consumables.
- These integrated devices facilitate easier monitoring, potentially reducing thrombotic events in hemodialysis patients.
Related Concept Videos
Dialysis
Dialysis
Acute kidney injury develops suddenly and can be caused by pre-renal causes (e.g., hypovolemia, shock), intrinsic renal causes (e.g., acute tubular necrosis), or post-renal causes (e.g., urinary obstruction). In contrast, chronic renal failure progresses gradually over time and is often...
Hemodialysis I: Introduction
Peritoneal Dialysis I: Introduction and Procedure
Hemodialysis III: Nursing Management
Continuous Renal Replacement Therapy