Related Experiment Video
Updated: May 7, 2026

06:38
Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Wound anatomy after type 1 Boston KPro using oversized back plates
Andrea Cruzat1, Anita Shukla, Claes H Dohlman
1Cornea and Refractive Surgery Service, Massachusetts Eye and Ear Infirmary and Department of Ophthalmology, Harvard Medical School, Boston, MA.
Cornea
|October 8, 2013
Summary
Oversized Boston Keratoprosthesis (KPro) back plates improve graft-host junction apposition and reduce retroprosthetic membrane (RPM) formation. This technique offers better wound healing and may prevent angle closure without adverse effects.
Area of Science:
- Ophthalmology
- Biomaterials Science
- Surgical Innovation
Background:
- Boston Keratoprosthesis (KPro) is crucial for vision restoration in severe corneal diseases.
- Graft-host junction integrity and anterior chamber angle health are critical for KPro success.
- Optimizing KPro back plate size may enhance surgical outcomes and reduce complications.
Purpose of the Study:
- To compare graft-host junction anatomy and anterior chamber angle after KPro implantation using oversized (9.5-mm) versus standard (8.5-mm) back plates.
- To evaluate the impact of back plate size on wound healing and retroprosthetic membrane (RPM) formation.
Main Methods:
- Anterior segment optical coherence tomography (AS-OCT) was used to image 16 patients (6 with 9.5-mm, 10 with 8.5-mm back plates) 6-12 months post-KPro.
- Assessment included graft-host junction location, corneal thickness, and anterior chamber angle.
- Clinical outcomes and RPM incidence were retrospectively reviewed.
Main Results:
- Oversized back plates ensured complete graft-host junction coverage and better apposition, reducing junctional wound thickness.
- Standard back plates did not fully cover the graft-host junction, leading to significantly thicker junctions.
- No significant RPM developed in the oversized group; one patient with a larger back plate experienced corneal melt at the KPro stem.
Conclusions:
- Oversized KPro back plates provide superior graft-host junction coverage without negatively impacting angle anatomy or wound healing.
- This approach may reduce RPM formation and prevent angle closure due to iris synechiae.
- Oversized back plates represent a promising strategy for improving KPro surgical outcomes.