Adaptive iterative dose reduction 3D versus filtered back projection in CT: evaluation of image quality
Raoul M S Joemai1, Wouter J H Veldkamp, Lucia J M Kroft
11 Department of Radiology, Leiden University Medical Center, C2-S, PO Box 9600, 2300RC Leiden, The Netherlands.
Objective:
The purpose of this study was to evaluate image quality with filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR 3D).
Materials And Methods:
Phantom acquisitions were performed at six dose levels to assess spatial resolution, noise, and low-contrast detectability (LCD). Spatial resolution was assessed with the modulation transfer function at high and low contrast levels. Noise power spectrum and SD of attenuation were assessed. LCD was calculated with a mathematic model observer applied to phantom CT images. The subjective image quality of clinical CT scans was assessed by five radiologists.
Results:
Compared with FBP, AIDR 3D resulted in substantial noise reduction at all frequencies with a similar shape of the noise power spectrum. Spatial resolution was similar for AIDR 3D and FBP. LCD improved with AIDR 3D, which was associated with a potential average dose reduction of 36% (range, 9-86%). The observer study showed that overall image quality improved and artifacts decreased with AIDR 3D.
Conclusion:
AIDR 3D performs better than FBP with regard to noise and LCD, resulting in better image quality, and performs similarly with respect to spatial resolution. The evaluation of image quality of clinical CT scans was consistent with the objective assessment of image quality with a phantom. The amount of dose reduction should be investigated for each clinical indication in studies with larger numbers of patients.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...