Synthesis of Soft Polysiloxane-urea Elastomers for Intraocular Lens Application
Published on: March 8, 2019
13.7K
Intraocular polyimide intraocular lens haptic breakage long-term postoperatively.
Shannon Stallings1, Liliana Werner1, Arturo Chayet1
1From the John A. Moran Eye Center (Stallings, Werner, Peck, Mamalis), University of Utah, Salt Lake City, Utah, and Advanced Vision Care (Masket), Los Angeles, California, USA; the Codet Vision Institute (Chayet, Camacho), Tijuana, Mexico.
Journal of Cataract and Refractive Surgery
|January 28, 2014
Summary
Two cases of silicone intraocular lenses (IOLs) with polyimide haptics showed spontaneous breakage after years of use. This brittleness occurred despite no signs of material degradation, highlighting a need for further research into polyimide biocompatibility.
Area of Science:
- Ophthalmology
- Biomaterials Science
Background:
- Presents two cases of three-piece silicone intraocular lenses (IOLs) featuring polyimide haptics.
- Highlights the importance of long-term biocompatibility studies for intraocular lens materials.
Observation:
- In the first case, an IOL implanted after posterior capsule rupture was explanted due to bullous keratopathy; its polyimide haptics shattered during manipulation.
- In the second case, an IOL initially implanted in-the-bag dislocated 15 years later due to spontaneous haptic breakage.
- Both explanted IOLs exhibited brittle haptics that fractured further upon manipulation for microscopy.
Findings:
- Scanning electron microscopy revealed no degradation of the polyimide haptics.
- Microscopic examination showed smooth surfaces with jagged fracture edges, indicating brittle failure.
- The observed haptic breakage occurred in situ without apparent surface degradation.
Implications:
- Suggests a potential long-term issue with polyimide haptic material leading to brittle fracture in silicone IOLs.
- Underscores the need for further investigation into the long-term performance and biocompatibility of polyimide in intraocular devices.
- May inform future IOL material selection and design to prevent late-stage complications.